PillCam Capsule Endoscopy
204
PillCam Recorder Maintenance
4.
To check the status of the PillCam recorder before an examination, remove it from the cradle and push 
the button on the back.
Manual Discharge 
If the cradle detects that the battery needs refreshing, it automatically discharges the battery before 
recharging it. The orange LED on the cradle blinks during discharging. Discharging is an overnight 
process that may take up to 12 hours.
We recommend that you manually discharge the battery every three months, even if it is not used. 
To discharge the battery:
1.
Open RAPID.
2.
Make sure the appropriate battery is inside the PillCam recorder.
3.
Insert the PillCam recorder into the cradle.
4.
From the 
Procedures
 screen, select the relevant PillCam recorder by clicking the 
Recorder
 bar.
The buttons on the right side of the screen become available.
5.
Click 
 to open the 
Recorder Information
 screen.
6.
At the bottom of the screen, click 
Start Discharge
.
A message appears: 
Recorder discharge may take more than 12 hours. Do you wish to continue?
7.
Click 
OK
.
While the battery is being discharged, its battery status indicates 
Discharging
:
In the bottom left corner of the 
Recorder Information
 screen.
In the PillCam recorder DR2 bar in the 
Recorders
 screen.
The orange LED on the cradle blinks.
8.
To return to other RAPID functions, click 
Close
.
9.
If you need to stop the discharge (also for automatic discharge) while it is in progress, return to the 
Recorder Information
 screen and click 
Cancel Discharge
.
If you stop the automatic discharge process in the middle, the battery LEDs may not indicate the 
correct battery status.
֠
֠
֠
֠
Note
Do not charge or discharge the battery in the vicinity of the patient.