Technical Description
Guidance and Manufacturer's Declarations
231
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is 
affected by absorption and reflection from structures, objects and people.
Note 3:
P is the maximum output power rating of the transmitter in watts (W) according to the 
transmitter manufacturer and d is the recommended separation distance in meters 
(m). 
Note 4:
Field strengths from fixed RF transmitters, as determined by an electromagnetic site 
survey
a
, should be less than the compliance level in each frequency range
b
.
Note 5:
Interference may occur in the vicinity of equipment marked with the following 
symbol: 
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones 
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be 
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF 
transmitters, an electromagnetic site survey should be considered. If the measured field strength in 
the location in which the PillCam recorder DR2 is used exceeds the applicable RF compliance level 
above, the PillCam recorder DR2 should be observed to verify normal operation. If abnormal 
performance is observed, additional measures may be necessary, such as re-orienting or relocating 
the PillCam recorder DR2.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications 
equipment and the PillCam recorder DR2
The PillCam recorder DR2 is intended for use in an electromagnetic environment in which 
radiated RF disturbances are controlled. The customer or the user of the PillCam recorder DR2 
can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF communications equipment (transmitters) and the PillCam recorder 
DR2 as recommended below, according to the maximum output power of the communications 
equipment.
Rated maximum 
output power of 
transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2,5 GHz
d = 2.3 P
0.01
Not applicable
0.12
0.23
0.1
Not applicable
0.38
0.73
1
Not applicable
1.2
2.3
10
Not applicable
3.8
7.3
100
Not applicable
12
23
For transmitters rated at a maximum output power not listed above, the recommended 
separation distance d in meters (m) can be determined using the equation applicable to the 
frequency of the transmitter, where P is the maximum output power rating of the transmitter in 
watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration - electronic immunity