PillCam Capsule Endoscopy
38
Preparing the Required Equipment
h.
Instruct the patient to contact the medical staff if one of the adhesive sleeves detaches from the 
patient's body.
i.
For procedures involving visualization of the colon: use the Regimen Manager to create and 
select pre- and post-ingestion instructions (see Regimen Manager on page 183); print for the 
patient and explain.
3.
Obtain the patient’s informed consent.
4.
Before the procedure, remind the patient about how to prepare for the procedure. Remind them about 
appropriate clothing:
If using the sensor belt, upper garment of thin, natural fiber cloth that is long enough to reach at 
least to hip level and will not ride up above the belt.
If using the sensor array, loose-fitting, two-piece opaque clothing. 
Preparing the Required Equipment 
Before the patient arrives for the procedure, verify that the following equipment and accessories are 
available:
RAPID (installed, configured, and open after accepting the license agreement)
PillCam capsule (ESO/UGI, SB, COLON)
PillCam recorder 
PillCam recorder pouch with shoulder strap
Printed instructions for the patient 
Prepared PillCam sensor belt or PillCam sensor array with sensors already inserted in sleeves 
Sensor Location Guide (ESO/UGI, SB, SB Pediatric, or COLON) if you are using a sensor array
Water and drinking cup
Any medication prescribed for the patient during the procedure 
Connecting the PillCam Recorder to RAPID for Check-in
The PillCam recorder must be connected to RAPID during patient check-in. The check-in process allows 
entering patient and procedure data for the procedure and saves the correct patient and procedure 
information to the PillCam recorder. The process is explained in detail in this section.
The PillCam recorder connects to the RAPID workstation or Personal Computer (PC) via the cradle.
1.
Make sure RAPID is On.
2.
Make sure the PillCam recorder is firmly inserted into the cradle. 
!
Caution
Connection of the PillCam recorder to the RAPID software for patient check-
in, video download, and recorder upgrade must be done using the cradle.