Performing PillCam Capsule Endoscopy
Real-Time Viewing with PillCam Recorder DR2 Only— External Mode
87
When in Real-Time viewing the PillCam recorder screen turns off and returns to the main screen after:
2 minutes of inactivity for colon visualization procedures/SB procedures,
30 minutes of inactivity for ESO/UGI procedures.
Real-Time Viewing with PillCam Recorder DR2 Only—
External Mode
To activate real-time viewing with external RAPID Real-Time viewer, connect either PillCam recorder 
model you are using to the external real-time viewer through the respective USB cable and activate the 
RAPID Real-Time viewer. For instructions on the operation of the RAPID Real-Time viewer, see the 
user manual for RAPID Real-Time viewer.
Removing Equipment from the Patient
If the patient has not already removed equipment, do so as follows:
1.
Disconnect the sensors from the PillCam recorder.
2.
Remove the PillCam recorder from the patient.
3.
Remove the sensors from the patient.
4.
Remove the PillCam recorder from the pouch and place it in the cradle.
5.
After each procedure, make sure to clean the equipment (see PillCam Sensor Cleaning on 
page 205)
 and charge the PillCam recorder (see Charging on page 203).
֠
֠
֠
֠
Note
AFR should be activated only after visually confirming entry into the small 
bowel