Given Imaging Limited CAPSH-3 Manual De Usuario

Descargar
Página de 250
Troubleshooting
Error Messages
211
Error Messages
For LED behavior, see LED Display on page 68. For error messages displayed on the PillCam recorder 
screen, see Error Messages on page 73.
Cancelling Data Copy 
displays error message that 
Video Copy was 
unsuccessful
Cancelling the Data Copy 
in mid-sequence triggers 
Video Copy message 
Message
 Solution
Full name not defined for this 
user. 
Please contact your IT system 
administrator.
In order to enable electronic signatures in the 
CE report, the full name field (full name will 
appear in electronic signature) must be 
defined in the user's relevant Windows 
account.
Electronic signature failed. 
Please contact your IT system 
administrator.
In order to enable electronic signatures in CE 
report, the user's relevant Windows account 
must be enabled for remote connection login. 
Possible reasons for this error message 
include:
Password Expired
Account Restrictions
Invalid Logon Hours
Account Lockout (or expired or disabled)
Password must be changed.
You do not have permission to 
access the Regimen Manager. 
Please contact your system 
administrator.
In order to allow physicians to edit and 
approve their regimens, the following must be 
done:
In the Settings screen, under the Other 
tab, define an existing folder as the 
Regimens Directory
This folder must be open for read and write 
permissions for the relevant users.
Patient Check-in procedure 
failed. Reason: Failed to update 
recorder software version.
Turn the recorder off and then on before the 
next check-in attempt. If the upgrade fails 
again, contact customer support. 
Problem
 Cause
Action