PillCam Capsule Endoscopy
226
Guidance and Manufacturer's Declarations
Guidance and manufacturer’s declaration - electronic immunity
The PillCam capsules are intended for use in the electromagnetic environment specified below. 
The customer or the user of the PillCam capsules should assure that it is used in such an 
environment.
Immunity test
IEC 60601 test level
Compliance 
level
Electromagnetic environment - 
guidance
Electrostatic 
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be at 
least 30%.
Electrical fast 
transient/burst
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/output 
lines
Not applicable Not applicable
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable Not applicable
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines
IEC 61000-4-11
<5% U
T
 
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
 
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
 
(>95% dip in U
T
)
for 5 sec
Not applicable Not applicable
Power frequency 
(50/60 Hz) magnetic 
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic 
of a typical location in a typical 
commercial or hospital 
environment.
NOTE: U
T
   is the AC mains voltage prior to application of the test level.