Given Imaging Limited CAPSH-3 Manual De Usuario
Preparing for PillCam Capsule Endoscopy
Preparing the Required Equipment
39
3.
Click
Patient Check-in
on the RAPID Home screen.
4.
In the
Procedures
screen, select the
Recorders
tab.
5.
Click the Recorder
bar to connect the function buttons to the PillCam recorder:
֠
֠
֠
֠
Note
•
•
When the DR3 battery approaches 400 cycles, RAPID will display a
warning. Contact customer service for a replacement battery when this
battery warning appears, or when the battery reaches 3 years of service,
whichever occurs first.
warning. Contact customer service for a replacement battery when this
battery warning appears, or when the battery reaches 3 years of service,
whichever occurs first.
•
Do not remove the PillCam recorder from its cradle before the end of the
check-in process.
check-in process.
Button
Action
Identifies the PillCam recorder for the selected recorder bar.
When you click this button, all the LEDs on the associated
PillCam recorder and cradle blink.
Prepares the PillCam recorder for the capsule endoscopy
procedure by entering patient data.
When you click this button, all the LEDs on the associated
PillCam recorder and cradle blink.
Prepares the PillCam recorder for the capsule endoscopy
procedure by entering patient data.