The Data Recorder LED starts blinking green.
5. Instruct the patient to take the M2A™ Capsule out of its
holder.
6. Ask the patient to put the M2A™ Capsule in his/her
mouth, and to hold it under the tongue WITHOUT
INGESTING.
INGESTING.
7. If after a short adjustment period, with the M2A™
Capsule in the patient’s mouth, the green light keeps on
blinking, ask the patient to ingest the capsule with a sip
of water.
blinking, ask the patient to ingest the capsule with a sip
of water.
Note:
The time elapsed between capsule activation and
ingestion should usually not exceed 1 minutes.
ingestion should usually not exceed 1 minutes.
In
order to minimize the potential for
radiofrequency interference from
the device after removing it from
the blister package and until it is
ingested by the patient, the M2A™
capsule should be tested as quickly
as possible. After verification
that the M2A™ capsule is
operational – it should be
immediately ingested by the
patient.
radiofrequency interference from
the device after removing it from
the blister package and until it is
ingested by the patient, the M2A™
capsule should be tested as quickly
as possible. After verification
that the M2A™ capsule is
operational – it should be
immediately ingested by the
patient.
8. In the Event Form, write down the ingestion time of the
M2A™ Capsule.