Given Imaging Limited GIVENRECORDER Manual De Usuario

Descargar
Página de 80
Introduction
Chapter 1
9
DataRecorder 3 LED Indications
The following table describes the LED indicators and their status/color for each of the most 
common DataRecorder 3 events/status. 
LEDs
DataRecorder 3 Status
DataRecorder is initialized with patient data and ready to capture capsule signals.
DataRecorder shuts down if no capsule signals are received for more than 90 minutes.
Blinking
DataRecorder is receiving capsule signals before capsule pairing is achieved.
Blinking rate = capsule frame rate
Blinking
DataRecorder is receiving paired capsule signals.
Blinking rate = capsule frame rate.
Blink every 5 seconds
DataRecorder has stopped receiving capsule signals for more than 5 seconds.
DataRecorder has started downloading.
DataRecorder has stopped recording because the memory card is full. 
Blinking
There is an instruction on the DataRecorder screen.
DataRecorder is malfunctioning.
Blinking
DataRecorder detects capsule signal, but is not recording it. 
This is a malfunction. Check the SensorArray connection or have patient move to a 
different location.
The LEDS on the navigation buttons blink in blue once every 5 seconds when the 
DataRecorder is on, out of the cradle and the LCD screen is off.
Pressing any of the navigation buttons when the LCD screen is OFF will turn the LCD 
screen ON.