Maintenance
Chapter 6
65
We recommend manually discharging the DataRecorder 2 battery once 
every three months, even if the DataRecorder 2 is not used. This will 
prevent the DataRecorder 2 from discharging automatically at an 
inconvenient time, since the discharge is an overnight process that may 
take up to 12 hours.
3.
As soon as the DataRecorder 2 or its Battery Pack are fully charged, the green LED turns on, 
and the Orange LED turns off. Leave the DataRecorder 2 in its Cradle until the next 
examination.
Note
You can check the status of the DataRecorder 2 by pushing the button on the 
back of the DataRecorder 2 momentarily. All 4 LEDs must be lit before an 
examination.
Manual Discharge of DataRecorder 2 
If the Cradle detects that the battery needs refreshing (i.e., the battery gauge needs recalibration), 
it will automatically discharge the battery before recharging it. The orange LED on the cradle 
blinks during discharging.
We recommend manually discharging the DataRecorder 2 battery once every three months, even 
if the DataRecorder 2 is not used. This will prevent the DataRecorder 2 from discharging 
automatically at an inconvenient time, since the discharge is an overnight process that may take 
up to 12 hours.
To discharge the DataRecorder 2 Battery
1.
Make sure the appropriate battery is inside the DataRecorder 2.
2.
Insert the DataRecorder 2 into its cradle.
3.
From the 
Procedures
 screen, select the relevant DataRecorder 2 by clicking the 
DataRecorder 2 bar.
The buttons on the right side of the screen become available.
4.
Click 
 to open the 
DataRecorder Info
 screen.
5.
At the bottom of the screen, click 
Start Discharge
.
A message appears: 
Discharge may take up to 12 hours. Are you sure you want to start 
discharge?
6.
Click 
OK
.
While the battery is being discharged, its battery status indicates 
Discharging
:
Push button
LEDs on the
DataRecorder 2
(at the back)