DOC-0062-02
Testing for Interference
Power LED turns on
Battery LED blinks,
Scanner turns off
Replace all eight
batteries
No
Yes
Yes
No
Attach TesTag
Detection LED turns on
Make sure the Scanner operates at least:
y
5 meters away from operating monitors
y
30 cm away from large metallic objects
y
1 meter away from electronic or magnetic
cards/badges and remote controls with
RFID tags
The Scanner is ready for the M2A
Patency examination
Yes
Call Given Customer
Support
Yes
Repeat the test in a
different location until
you find a suitable
environment
No
Power LED and Battery
LED turn on
No
Yes
No