Appendix: Electrical Safety 
System Components 
SmartPill Capsule Pack 
Each capsule pack contains a single-use capsule, calibration buffer, instructions for use 
and a patient diary.  
The capsule measures pressure, pH, and temperature to determine transit times of the 
stomach, small bowel, and colon. Transit times derived by capsule motility procedures 
provide alternatives to other tests such as gastric emptying scintigraphy, whole gut 
scintigraphy, and radio-opaque markers. 
Caution 
Do not use a SmartPill capsule if it has been dropped as this may affect function. 
 
Capsule Operational Specifications 
Attribute 
Specification 
Pressure Accuracy 
 
0-99 mmHg     ± 5 mmHg 
100-350 mmHg      ± 10% of applied pressure 
Pressure Range 
0-350 mmHg 
pH Accuracy 
± 0.5 pH units 
pH Range 
1–9 pH units 
Temperature Accuracy 
± 1°C (between 20 – 42ºC) 
Transmission Frequency 
434.2 MHz (radiating between 426 – 445 MHz) 
Battery Life 
Capsule and Data Receiver, >5 days 
Weight 
4.5 grams 
Size 
26 x 13mm 
SmartBar 
The SmartBar is a standardized meal that is ingested immediately before capsule 
ingestion. To accurately measure gastric emptying patients must consume a standard 
meal immediately before ingesting the capsule.