Appendix: Electrical Safety 
Step 6: Enter Pressure Calibration Code 
1. Locate the calibration code found on the lid of the capsule container. Enter the 
code.  
 
2. Click Next >.  
Step 7: Add pH Calibration Buffer 
Caution 
Once exposed to buffer, the capsule must be ingested within 2 hours or discarded. Confirm that the patient is 
available and prepared to proceed with the test before you begin Step 7. 
1. Put gloves on and follow the prompts. 
Caution 
Wear gloves when handling the capsule. Do not touch the calibration buffer or capsule with your finger as 
this could cause an electrostatic charge.