Troubleshooting (page 2 of 2)
Electrodes must be securely attached to patient.
Patient should remain still while recording.
Replace Patient Cable. Pulling on lead wires may damage
cable.
Verify the recording did not take place near a source of
electromagnetic interference (fluorescent lights, computer
monitors, or household appliances).
Patient should remain still while recording.
Replace Patient Cable. Pulling on lead wires may damage
cable.
Verify the recording did not take place near a source of
electromagnetic interference (fluorescent lights, computer
monitors, or household appliances).
Noise artifact on recorded
ECG at patient location
ECG at patient location
Move electrodes slightly to the right or left of the original
location.
location.
Rising tone
Ready to record
Fusion Wireless Recorder
18
Service and Maintenance
Cleaning
Cleaning should occur before each patient use and more frequently if
needed.
needed.
Remove the batteries before cleaning the recorder. Clean the battery
terminals with a soft dry cloth. Dampen a soft cloth with mild detergent
and water to clean the recorder, lead wires, and belt clip.
terminals with a soft dry cloth. Dampen a soft cloth with mild detergent
and water to clean the recorder, lead wires, and belt clip.
Remove any adhesives from the patient lead wires with an adhesive tape
remover solution or swab. Use a mild disinfectant. Do not use alcohol
or acetone on the lead wires since they could stiffen and the insulating
plastic could crack.
remover solution or swab. Use a mild disinfectant. Do not use alcohol
or acetone on the lead wires since they could stiffen and the insulating
plastic could crack.
Service
If there is a problem with the Recorder, review the problem descriptions
and solutions listed on the next page. If additional assistance is required
contact customer support via phone, Fax or E-mail listed below. Call
customer support before returning a Recorder to make shipping
arrangements.
and solutions listed on the next page. If additional assistance is required
contact customer support via phone, Fax or E-mail listed below. Call
customer support before returning a Recorder to make shipping
arrangements.
A.
Note there isn’t any preventative inspection or maintenance that
can be performed by the end user.
can be performed by the end user.
Service Items and Accessories
Note: Only authorized accessories are
permitted.
permitted.
Description
Part Number
Patient Cable, 3 channel, 5 lead
350-0302-00
Patient Cable, 2 channel, 4 lead
350-0302-01
Patient Cable, 1 channel, 3 lead
350-0302-02
Patient Cable, 1 channel, 2 lead
350-0302-03
Patient Cable, 2 channel, 3 lead
350-0302-04
Recorder belt clip / Holster
100-1910-001
Operator manual
600-0645-00
AA Lithium Battery Pack
350-0294-00
USB 2.0 Data Cable
200-2792-001