The DELETE menus allows for deleting current data, all data, and group data. The following 
parameters are available in the DELETE Menu: 
Delete Current data: Deletes current reading and updates the statistics (AVG, MIN, MAX, etc.) 
Delete All data: Delete all reading and statistical data. 
Delete Group data: This function duplicates the “Delete all data” function with additional, 
deletions of High alarm, Low alarm, and one‐ and two‐point calibrations. 
1.  Press the LEFT button to access the programming menu. 
2.  User the DOWN ARROW button to scroll down to DELETE. 
3.  Press LEFT to open the DELETE function. 
4.  User the ARROW keys to scroll to CURRENT, ALL, or GROUP. 
5.  Press LEFT again to choose CURRENT, ALL, or GROUP. 
MEASUREMENT VIEW Menu  
The Measurement View menu allows for scrolling through the readings in all of the groups. 
1.  Press the LEFT button to access the programming menu. 
2.  User the DOWN ARROW button to scroll down to MEASUREMENT VIEW.  
3.  Press LEFT to open the MEASUREMENT VIEW parameter. 
4.  User the ARROW buttons to scroll through the stored readings.  
CALIBRATION Menu  
The Calibration menu allows the user to enable/disable the calibration utility. The Calibration 
menu also allows the user to delete Zero calibration data for both ferrous (Zero F) and non‐
ferrous (Zero N) modes. 
1.  Press the LEFT button to access the programming menu. 
2.  Use the DOWN button to scroll down to CALIBRATION.  
3.  Press LEFT to open the CALIBRATION parameter. 
4.  Press UP/DOWN to delete zero point of Fe or Non‐Fe. 
5.  Press LEFT button for OK 
6.  Press RIGHT button to return. 
Measurement Considerations 
1.  After calibration, measurements should meet the published accuracy specifications. 
2.  Strong magnetic fields can affect the readings. 
3.  When using the statistical analysis functions for obtaining a mean value, take several 
readings of the same measurement area. False readings or outliers can then be removed 
(deleted) using the programming menu. 
4.  The final reading is derived from a statistical calculation with regard to the meter’s 
published accuracy specifications.