EMBLEM™ S-ICD PROGRAMMER:  OPERATIOn
Figure 49: Capture all sense vectors
Once captured, the three S-ECGs can be viewed by following the steps outlined in Viewing 
Beeper Control
The pulse generator has an internal alert system that emits an audible beeping tone when an alert 
condition occurs. Alert conditions may include Elective Replacement Indicator (ERI), End of Life 
(EOL), electrode impedance out of range, prolonged charge times, failed Device Integrity Check and 
irregular battery depletion. 
Caution:  Patients should be advised to contact their physician immediately 
whenever they hear beeping tones coming from their device.
Once an alert condition is triggered, audible tones beep for 16 seconds every nine hours until the 
alert condition is resolved. Beeper Control  allows the beeper to be temporarily suspended (Reset 
Beeper) or permanently inactivated (Disable Beeper).
Note:  Access to the Beeper Control display screen is enabled only when an alert condition 
occurs. If an alert condition is activated, a notification screen will appear upon 
connection.