Q-TECH™ PROGRAMMER:  OPERATiOn
Performing an SQ-RX Follow-up
Sensing Configuration and Automatic Setup
It is not necessary to perform Automatic Setup at each follow-up. If Automatic Setup is performed, 
resulting in a vector change, sensing should be reevaluated. After the setup process is complete, 
evaluate the streaming S-ECG during a pectoral exercise. Sensing performance during high rate 
exercises can also be performed. Acceptable sensing will yield “S” markers synchronous to all QRS 
complexes. If other markers are noted, use the Manual Setup process to evaluate other sensing 
configurations.
Note:   If Manual Setup was previously used to override a sensing configuration, 
careful consideration should be taken when selecting Automatic Setup.
If an update to the reference S-ECG is desired due to a change in the patient’s resting ECG, follow 
the Acquire Reference S-ECG instructions.
Viewing SQ-RX Device Status
Once communication is established, the programmer displays the Device Status screen with 
information regarding the current episodes and battery status of the SQ-RX device.
To view this information:
1. 
Select the Main Menu icon.
2. 
Select the Follow Up button.
3. 
Select the Device Status icon on the Navigation Bar.
4. 
The Device Status screen will appear showing an overview of all SQ-RX device activity 
since the last communication session (Figure 35).