Bird Technologies Group 5PI613805 Manual De Usuario

Descargar
Página de 38
TXRX Systems Inc.                                Manual 7-9408-1.2                                 07/25/05                                  Page 23
61-38-05 UserMan page 23 of 38
either the transmission or reflection response
may be unavoidable.
Bandpass Filters
The bandpass filters pass one narrow band of fre-
quencies (the passband) and attenuate all others
with increasing attenuation above and below the
pass frequencies. The insertion loss setting deter-
mines the filters selectivity and maximum power
handling capability. Insertion loss is set at the fac-
tory.
Cavity tuning follows a two step process. First the
cavity is temporarily disconnected from the system
and rough tuned, this will ensure the response
curve is very close to it's ideal. Next the cavity is
reconnected to the system and fine tuned. This is
done in large systems such as the model 61-38-05
in order to eliminate any slight distortions which
could result from interactions with other cavities.
The pass frequency is the only field adjustable
parameter found in the individual Bandpass reso-
nant cavity filters. Adjustment of the tuning rod on
these filters will allow the passband to be centered
at the desired frequency. The insertion loss of each
cavity is not field adjustable.
REQUIRED EQUIPMENT
Due to the sensitivity of the adjustments, it is
strongly recommended that the proper equipment
be used when tuning the individual filters, other-
wise the filter should be sent to the factory or an
authorized representative for retuning. The follow-
ing equipment or it's equivalent is recommended in
order to properly perform the tuning adjustments.
1. IFR A-7550 spectrum analyzer with optional
tracking generator installed.
2. 7/16” wrench.
3. Double shielded coaxial cable test leads
(RG142 B/U or RG223/U).
4. Female union (UG29-N or UG914-BNC).
TUNING PROCEDURE
The following is an outline of the general proce-
dure.
1. Turn off the system power and disconnect the
cables that are attached to the cavity. 
2. Setup the analyzer / generator for the desired
frequency and bandwidth (center of display)
and also a vertical scale of 2 dB/div. Set the
sweep width of the display to 100 KHz.
4. A zero reference must first be established at the
IFR A-7550 before making measurements. This
is accomplished by temporarily placing a
"female union" between the generator output
and the analyzer input.
5. The flat line across the screen is the generator's
output with no attenuation, this value will
become our reference value by selecting the
4 Section 
Helical Preselector
Analyzer
Input
Generate
Output
+30
+40
+20
+10
0
-10
-20
-30
-40
Bridge
50 ohm Load
Figure 12: Observing preselector return loss.