8. MAINTENANCE 
 
8.1 DAILY MAINTENANCE 
Use the peel-off clear covers for the laser console supplied with the system. Use disinfectant to 
wipe down the front panel and handpiece holder of the EPIC system after each procedure.     
Do not use bleach or abrasive cleansers. 
8.2 CLEANING AND STERILIZATION PROCEDURES 
The contamination control suggested for the EPIC surgical handpiece and tips is the steam 
sterilization method. However, before sterilization, the EPIC reusable handpiece should be 
carefully cleaned per the following procedure. 
CAUTION: 
Tips are single-use only to avoid cross-contamination and are designed to 
withstand a single sterilization cycle; they must be disposed of after use in a 
biohazard medical waste Sharps container. 
 
Handpieces are reusable and must be cleaned and sterilized between 
patients to avoid cross-contamination. 
 
Cleaning and Disinfecting Instructions for the Surgical Handpiece, and the 
Reusable Fiber Optic Cable 
The cleaning process is intended to remove blood, protein and other potential contaminants 
from the surfaces and crevices of reusable accessories. This process may also reduce the 
quantity of particles, microorganisms and pathogens present. Cleaning should be performed 
prior to sterilization and must be conducted only by qualified office personnel trained to perform 
the procedure and handle the EPIC fiber optic delivery system. 
 Wear protective latex gloves when handling the contaminated delivery system. 
To disinfect the fiber cable, wipe the entire cable, including the shaft, with an appropriate 
disinfecting solution, such as Cavicide™ or a similar quaternary ammonium compound 
product (containing 20% alcohol or less), and follow the manufacturer’s instructions.  Avoid 
getting any liquid or debris near the distal end of the fiber cable. 
 
 
 
WARNING: 
Changes or modifications not expressly approved by Biolase could 
void the user’s authority to operate the equipment.