Badger Meter Inc RANG312 Manual De Usuario

Descargar
Página de 64
Page 13
1-12
Installation & Operation Manual
Charging the Handheld Battery
The handheld should arrive fully charged. However, charging the battery before use is a good practice. 
The battery pack assembly includes the batteries to power the unit, the power port used to charge the unit and the I/O ports 
that are used to communicate with other devices using a USB or serial cable.
The AC charger that ships with the unit has adapters to fit almost every electrical outlet. Connect the appropriate outlet 
adapter to the AC charger. Plug the charger into an electrical outlet. Make sure the handheld is fully seated in the charger 
and the charging pin is fully engaged. 
The handheld is fully charged in 
approximately 4.5 hours. Acceptable 
temperature range for charging the 
battery is between +5° C and +35° C (+41° 
F and +95° F).
The LED (light emitting diode) on the top 
left corner of the handheld displays the 
charging status as shown in this table.
Normal usage
Off 
No charger present or unit is unplugged
Yellow
Charging
Green
Charging complete
Error conditions
1 - 2 short flashes * 
Temperature out of range for charging
3 short flashes 
Wrong charger
Note: If the fully-charged unit is still connected to the charger, SettingsSystemPower status shows the battery 
power remaining as if the unit were still on battery power.
 Always use the charger included with your unit. Use of other chargers will void the warranty.
Conserving the Battery
The handheld conserves battery power by turning itself off when not in use. 
Note: When the Read screen or Quick Read screen is in use, the handheld will NOT turn off automatically because the 
ORION reading functions are active. If you wish to conserve the battery, be sure to return to the Main Menu so the 
automatic off function is active. 
Press and release the green power key on the keypad to turn the unit back on. The display shows the same screen that was 
displayed when the unit turned off. 
Battery Life
Wireless
The handheld has been designed to operate with the internal BlueTooth® and 802.11 wireless communication features and 
will automatically turn them off when not in use. Do not turn on the wireless features unless instructed. Leaving the wireless 
options on reduces battery life.
Charging Cradle
When using the handheld cradle to charge the battery pack, be sure the handheld is fully seated into the cradle. The cradle 
is designed to be set up on a flat surface or mounted on a wall. Up to 10 cradles can be connected together and use a single 
power cord. 
To verify the unit is being charged, the top line of the screen will indicate (Batt: A/C). It may take a few minutes for the screen 
to update from the rate of battery charge to the indication that it is being charged by A/C power. To immediately check the 
battery pack charging status, exit the ORION Field Application and then restart. The top line of the screen shows the battery 
charge status.