5.1.4  Guidance and manufacturer’s declaration – electromagnetic immunity 
The model SHC-100 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the model SHC-100 should assure that it is used in such an environment. 
Immunity test 
IEC 60601 test level 
Compliance 
resoult 
Electromagnetic 
environment – guidance 
Conducted RF  
IEC 61000-4-6  
3 Vrms 
Pass  
Portable and mobile RF 
communications equipment should be 
used no closer to any part of the model 
SHC-100, including cables, than the 
recommended separation distance 
calculated from the equation applicable 
to the frequency of the transmitter. 
 
Recommended separation distance 
d = 1.2
 
d = 1.2 
80 MHz to 800 MHz 
d = 2.3 
800 MHz to 2,5 GHz 
 
Where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the recommended 
separation distance in meters (m). 
 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey
α
, should are 
less than the compliance level in each 
frequency range
β
.  
 
Interference may occur in the vicinity 
of equipment marked with the 
following symbol: 
 
 
 
50 kHz to 80 MHz 
6Vrms in ISM bands 
between 150 kHz 
and 80 MHz 
180% AM at 1 kHz 
Radiated  
RF EM 
IEC 61000-4-3 
3 V/m  
80 MHz to 2.7 GHz 
80% AM at 1 kHz 
Pass 
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.  
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects, and people.
 
α:
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 
radios, amateur radio, AM, and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If 
the measured field strength in the location in which the model SHC-100 is used exceeds the applicable RF compliance level 
above, the model SHC-100 should be observed to verify normal operation. If abnormal performance is observed, additional 
measures may be necessary, such as reorienting or relocating the model SHC-100.  
β: 
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.