There are three general categories of questionable arrhythmia calls.
1. False Positive Calls result when the arrhythmia system signals an arrhythmia 
when there is none. It is important to understand how the algorithm analyzed 
the rhythm.
2. False Negative Calls result when the arrhythmia system fails to detect and 
signal an arrhythmia event when one is present.
3. Incorrect Analysis occurs when the system signals that an arrhythmia event 
occurred, but classifies it incorrectly. Note that it is important that the system 
alarmed and alerted the system operator to a change in the patient rhythm.
If there are too many false positive calls, then the system operator may become desen-
sitized to the alarms. That is why it is important that you understand what factors con-
tribute to false positive calls and correct this situation when possible.