Panorama™ Operating Instructions
0070-10-0690-01
5 - 21
Patient Setup Functions
Demographics Tab
The Retrieve Discharged 
Patient Settings
 tab was 
closed while the bedside 
device’s Patient Menu is active.
The Arrhythmia Menu dialog 
box was closed while the 
bedside device’s Patient Menu 
is active.
The Pacer Settings dialog box 
was closed while the bedside 
device’s Patient Menu is active.
The CO Setup Menu is opened 
at the bedside monitor while 
the Demographics tab is 
active at the central station or 
WorkStation.
The Demographics tab is 
active at the central station or 
WorkStation and then the CO 
Setup Menu is opened at the 
bedside monitor. 
The Hemodynamic 
Calculations Menu is opened 
at the bedside monitor while 
the Demographics tab is 
active at the central station or 
WorkStation.
The Demographics tab is 
active at the central station or 
WorkStation and then the 
Hemodynamic Calculations 
Menu is opened at the bedside 
monitor.
The Arrhythmia Menu at the 
central station was closed and 
all changes were discarded.
The Arrhythmia Menu was 
opened at the bedside monitor.
Use the bedside monitor to 
make the arrhythmia changes. 
Close the Arrhythmia Menu at 
the bedside monitor and then 
reenter the Arrhythmia Menu 
dialog box at the central 
station or WorkStation.
MESSAGE/ISSUE *
REASON
SOLUTION
*
Messages are shown in all bold text.