Panorama™ Operating Instructions
0070-10-0690-01
xv
Warnings, Precautions and Notes
Introduction
CAUTION:
The Telepack with Nonin SpO
2
 module has motion tolerant 
software that minimizes the likelihood of motion artifact 
being misinterpreted as good pulse quality. In some 
circumstances, however, the Telepack with Nonin SpO
2
 
module may still interpret motion as good pulse quality. This 
applies to all available outputs (i.e. SpO
2
, HR, Pleth, PPG).
CAUTION:
A functional tester cannot be used to assess the accuracy of 
the pulse oximeter probe or a pulse oximeter monitor. 
CAUTION:
The Telepack with Nonin SpO
2
 module complies with 
International Standard IEC 60601-1-2:2005 for 
electromagnetic compatibility for medical electrical 
equipment and/or systems. This standard is designed to 
provide reasonable protection against harmful interference 
in a typical medical installation. However, because of the 
proliferation of radio-frequency transmitting equipment and 
other sources of electrical noise in healthcare and other 
environments, it is possible that high levels of such 
interference due to close proximity or strength of a source 
might disrupt the performance of this device. Medical 
electrical equipment needs special precautions regarding 
EMC, and all equipment must be installed and put into 
service according to the EMC information specified in this 
manual.
CAUTION:
Datascope requires that the designated cleaning plug
(P/N 0003-00-0055)
 be used when cleaning and 
disinfecting the Nonin SpO
2
 module. If fluids enter the 
device, it may affect the functionality of the module or 
render it unusable. 
CAUTION:
In compliance with the European Directive on Waste 
Electrical and Electronic Equipment (WEEE) 2002/96/EC, do 
not dispose of this product as unsorted municipal waste.
CAUTION:
It is important to notify Datascope Service personnel when 
the system event message “System Service required - 
Contact Datascope or Authorized Datascope Distributor 
(300).” is displayed. This message indicates that while data 
was being written to or read from the system hard drive, an 
error was detected. Until service personnel correct this issue, 
patient data may be lost (for example, demographic 
information, alarm configurations, and historical patient 
data). Real time parameter monitoring and alarm checking 
will not be affected by this message. This message will be 
displayed until reset by service personnel.