Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Trends Tab
Views
7 - 18
0070-10-0690-01
Panorama™ Operating Instructions
No more data found
(Previous Event Button)
There is no event data in the 
Trend database.
Wait until an event is recorded to the 
Trend database.
None of the rows that are 
below the currently 
highlighted row contain a 
parameter with an event 
condition.
Select the OK button to close the 
message box and return to the List 
view of the Trends tab.
The last row in the list is 
already highlighted.
Select the OK button to close the 
message box and return to the List 
view of the Trends tab.
No more events found
(Next Event Button)
There is no event data in the 
Trend database.
Wait until an event is recorded to the 
Trend database.
None of the rows that are 
above the currently 
highlighted row contain a 
parameter with an event 
condition.
Select the OK button to close the 
message box and return to the List 
view of the Trends tab.
The first row in the list is 
already highlighted.
Select the OK button to close the 
message box and return to the List 
view of the Trends tab.
The Graphic and Trend 
Display 
sidebar buttons 
are disabled.
There is no data in the Trend 
database because the 
Trends tab was selected 
before the first interval was 
reached or before the first 
NIBP measurement occurred.
Wait until the first interval is reached 
or until the first NIBP measurement 
occurs.
A patient has not been 
admitted to the selected tile.
Admit a patient to the selected tile or 
choose a tile that already has an 
admitted patient.
The Trends List data is still 
loading.
Wait for the Trends List data to finish 
loading.
The Print Selection dialog 
box is open.
Close the Print Selection dialog 
box.
The Print menu bar button 
is disabled.
The trends data is loading.
Wait until the trends data is loaded.
MESSAGE/ISSUE *
REASON
SOLUTION
*
Messages are shown in all bold text.