Trends Tab
Views
7 - 22
0070-10-0690-01
Panorama™ Operating Instructions
7.2.4.1
Troubleshooting
This section lists some of the potential messages and issues that may occur while using the 
Graphic view of the Trends tab.  
MESSAGE/ISSUE *
REASON
SOLUTION
There is no trend data in 
the display.
A patient has not been 
admitted to the selected tile.
Admit a patient to the selected tile or 
choose a tile that already has an 
admitted patient.
The Trends tab was selected 
before the first interval was 
reached or before the first 
NIBP measurement occurred.
Wait until the first interval is reached 
or until the first NIBP measurement 
occurs.
The parameter areas only 
display the parameter 
label and/or the unit of 
measurement.
There is no data at the 
location of the cursor line.
Move the cursor line to a point on 
the time line where data is known to 
exist.
No more data found
(Page Left Button)
There is no data in the Trend 
database because the 
Trends tab was selected 
before the first interval was 
reached or before the first 
NIBP measurement occurred.
Wait until the first interval is reached 
or until the first NIBP measurement 
occurs.
The first recorded data has 
been reached. No further 
data exists before this time.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
No more data found
(Page Right Button)
There is no data in the Trend 
database because the 
Trends tab was selected 
before the first interval was 
reached or before the first 
NIBP measurement occurred.
Wait until the first interval is reached 
or until the first NIBP measurement 
occurs.
The most recent recorded data 
has been reached. No further 
data exists after this time.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
Time out of range
(Zoom In Button)
The minimum display time 
interval has been reached.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
Time out of range
(Zoom Out Button)
The maximum display time 
interval has been reached.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
Time out of range
(Left Arrow Button)
The minimum display time 
interval has been reached.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
Time out of range
(Right Arrow Button)
The maximum display time 
interval has been reached.
Select the OK button to close the 
message box and return to the 
Graphic view of the Trends tab.
The List sidebar button is 
disabled.
A message box is being 
displayed.
Close the message box.
*
Messages are shown in all bold text.