Patient Reports Tab
Report Functions
8 - 28
0070-10-0690-01
Panorama™ Operating Instructions
A WMTS 608 wireless bedside 
device is in use. 
No solution required. 
The selected patient tile is 
relearning phase, or is in a 
Communications Lost or 
Standby state.
Wait for the patient tile to 
resume monitoring.
The ST Report button is 
hidden. 
The selected patient tile is in a 
Waiting for Patient state.
Activate the destination device.
The Print All Strips report 
button is disabled. 
The selected patient tile is in a 
Communications Lost or 
Standby state.
Wait for the patient tile to 
resume monitoring.
The Print All Strips button is 
hidden.
The selected patient tile is in a 
Waiting for Patient state.
Activate the destination device.
The report will not print
There are several reasons that 
a report would not print. There 
could be a communication 
problem between the printer 
and the Panorama Central 
Station and/or there could be 
a problem with the printer.
If there is an error that appears 
such as Printer out of 
Paper
 you will need to 
remedy that situation. 
If the No Printer connected 
or Printer not Available
 
error is displayed, verify that 
the printer was properly 
configured with the Panorama 
Central Station. Refer to the 
Panorama Printer 
Configuration Manual (P/N 
0070-00-0561).
The Delta T parameter does not 
display in the All Strips report.
One of the two temperature 
sources used to calculate the 
Delta T value is not valid.
Verify that the two temperature 
sources have been properly 
configured.
Page On Alarm column not 
displayed in Patient Alarm 
Report
Paging option is not enabled at 
the central station.
If available, enable Paging 
option. If feature unavailable, 
no solution required. 
All Arrhythmia On/Off setting 
set to Off
Arrhythmia is disabled for a 
selected patient tile.
If available, enable arrhythmia 
for the selected patient tile. If 
feature or license is 
unavailable, no solution 
required. 
In the Patient Alarm Report’s 
Arrhythmia Settings section, a
“---” is displayed as the value 
for Asystole Delay, V-Tach 
Threshold, and V-Tach Rate 
Arrhythmia is disabled for a 
selected patient tile.
If available, enable arrhythmia 
for the selected patient tile. If 
feature or license is 
unavailable, no solution 
required. 
Paging Delay Settings section 
not displayed in the Patient 
Alarm Report. 
This section will only be 
displayed when Paging is 
enabled at the central station.
If available, enable Paging 
option. If feature unavailable, 
no solution required. 
MESSAGE/ISSUE * 
REASON
SOLUTION
*
Messages are shown in all bold text.