Equipment Setup Tab
System Setup Functions
9 - 38
0070-10-0690-01
Panorama™ Operating Instructions
Suffix of Device ID last 
character must be ‘x’ or 
‘X’.
 
This message appears when 
the incorrect device suffix is 
entered in the Device ID box.
Select the OK button to close 
the dialog box. Enter the 
correct Device ID suffix.
This Device ID already 
exists!
 
This message appears when a 
duplicate Device ID is entered 
in the Device ID field.
Select the OK button to close 
the dialog box. Enter a new 
Device ID in the Device ID 
field.
You have reached the 
maximum number of 
channel licenses 
available.
 
This message appears when all 
available channels on the 
Panorama Central Station are 
occupied.
Select the OK button to close 
the dialog box. Delete a piece 
of equipment from an existing 
channel if it is no longer in use. 
The system administrator can 
also purchase additional 
channel licenses.
This version of software 
does not support more 
than 16 monitored 
patients.
 
This message appears when 
the maximum number of 
patients is being monitored on 
the Panorama Central Station.
Select the OK button to close 
the dialog box. Discharge a 
patient if the patient is no 
longer being monitored. 
System Setup menu button is 
disabled.
The System Setup function is 
already open on another 
Panorama display and cannot 
be opened on multiple 
displays.
Close the System Setup 
function on the Panorama 
display where it is open.
Cannot delete a device 
connected to a patient
 
The device cannot be deleted 
because it is currently 
connected to a patient.
Select the OK button to close 
the dialog box. Disconnect the 
patient from the selected 
device before attempting to 
delete it.
The Tile button, Type button 
and Device ID field are 
disabled.
The selected equipment cannot 
be edited because it is 
currently monitoring a patient.
This equipment cannot be 
edited until a patient discharge 
is performed.
Unable to perform 
operation on tile: Central 
to Central Transfer In in 
progress.
Equipment Setup tab Delete 
button was selected for 
equipment that was not 
actively monitoring a patient, 
while a Central to Central 
Transfer In is in progress to that 
same device.
Wait for the Central to Central 
Transfer to complete, then try 
again. 
MESSAGE/ISSUE * 
REASON
SOLUTION
*
Messages are shown in all bold text.