Mindray DS USA Inc. WT0707-00000 Manual De Usuario
Equipment Setup Tab
System Setup Functions
9 - 38
0070-10-0690-01
Panorama™ Operating Instructions
Suffix of Device ID last
character must be ‘x’ or
‘X’.
character must be ‘x’ or
‘X’.
This message appears when
the incorrect device suffix is
entered in the Device ID box.
the incorrect device suffix is
entered in the Device ID box.
Select the OK button to close
the dialog box. Enter the
correct Device ID suffix.
the dialog box. Enter the
correct Device ID suffix.
This Device ID already
exists!
exists!
This message appears when a
duplicate Device ID is entered
in the Device ID field.
duplicate Device ID is entered
in the Device ID field.
Select the OK button to close
the dialog box. Enter a new
Device ID in the Device ID
field.
the dialog box. Enter a new
Device ID in the Device ID
field.
You have reached the
maximum number of
channel licenses
available.
maximum number of
channel licenses
available.
This message appears when all
available channels on the
Panorama Central Station are
occupied.
available channels on the
Panorama Central Station are
occupied.
Select the OK button to close
the dialog box. Delete a piece
of equipment from an existing
channel if it is no longer in use.
The system administrator can
also purchase additional
channel licenses.
the dialog box. Delete a piece
of equipment from an existing
channel if it is no longer in use.
The system administrator can
also purchase additional
channel licenses.
This version of software
does not support more
than 16 monitored
patients.
does not support more
than 16 monitored
patients.
This message appears when
the maximum number of
patients is being monitored on
the Panorama Central Station.
the maximum number of
patients is being monitored on
the Panorama Central Station.
Select the OK button to close
the dialog box. Discharge a
patient if the patient is no
longer being monitored.
the dialog box. Discharge a
patient if the patient is no
longer being monitored.
System Setup menu button is
disabled.
disabled.
The System Setup function is
already open on another
Panorama display and cannot
be opened on multiple
displays.
already open on another
Panorama display and cannot
be opened on multiple
displays.
Close the System Setup
function on the Panorama
display where it is open.
function on the Panorama
display where it is open.
Cannot delete a device
connected to a patient
connected to a patient
The device cannot be deleted
because it is currently
connected to a patient.
because it is currently
connected to a patient.
Select the OK button to close
the dialog box. Disconnect the
patient from the selected
device before attempting to
delete it.
the dialog box. Disconnect the
patient from the selected
device before attempting to
delete it.
The Tile button, Type button
and Device ID field are
disabled.
and Device ID field are
disabled.
The selected equipment cannot
be edited because it is
currently monitoring a patient.
be edited because it is
currently monitoring a patient.
This equipment cannot be
edited until a patient discharge
is performed.
edited until a patient discharge
is performed.
Unable to perform
operation on tile: Central
to Central Transfer In in
progress.
operation on tile: Central
to Central Transfer In in
progress.
Equipment Setup tab Delete
button was selected for
equipment that was not
actively monitoring a patient,
while a Central to Central
Transfer In is in progress to that
same device.
button was selected for
equipment that was not
actively monitoring a patient,
while a Central to Central
Transfer In is in progress to that
same device.
Wait for the Central to Central
Transfer to complete, then try
again.
Transfer to complete, then try
again.
MESSAGE/ISSUE *
REASON
SOLUTION
*
Messages are shown in all bold text.