Wave Gain Tab (System)
System Setup Functions
9 - 60
0070-10-0690-01
Panorama™ Operating Instructions
NOTE:
Changing the ECG wave gain does not affect the arrhythmia 
analysis.
3. Select the Resp Gain button until the desired Resp Gain wave gain is displayed.
The settings available for the Resp Gain wave gain include 1, 2, 3, 4, and 5. The factory 
default setting is 1.
4. Select the CO2 (%, kPa, mmHG) button until the desired CO
2
 wave gain is 
displayed.
• The settings available for the CO
% wave gain include 5.0, 7.5, and 10.0 %. The 
factory default setting is 10.0 %.
• The settings available for the CO
2
 kPa wave gain include 5.0, 7.5, and 10.0 kPa. The 
factory default setting is 10.0 kPa.
• The settings available for the CO
2
 mmHg wave gain include 40, 60, and 100 mmHg. 
The factory default setting is 100 mmHg.
5. Select the O2 % button until the desired O
2
 wave gain is displayed.
The settings available for the O
2
 wave gain include 18 - 30, 18 - 60, and 18 - 100 %. 
The factory default setting is 18 - 100%.
6. Select the AGENT % button until the desired Agent wave gain is displayed.
The settings available for the agent wave gains include 0 - 1.0, 0 - 2.5, 0 - 5.0, 0-10.0, 
0-15.0, and 0-20.0 %. The factory default setting is 0-20.0%.
9.15.3
Sidebar Button - Normal Screen
Select the Normal Screen sidebar button to return the display to the full screen mode, and 
close all tabs, message boxes, and dialog boxes.