Mindray DS USA Inc. WT0707-00000 Manual De Usuario

Descargar
Página de 484
Parameter Specifications
Parameter Specifications
12 - 2
0070-10-0690-01
Panorama™ Operating Instructions
12.1
Parameter Specifications
The parameters displayed on the Panorama Central Station are dependent on the type of 
equipment connected to the Panorama system.
12.1.1
Heart Rate: HR
The Heart Rate parameter obtains data from ECG and IBP sources.
 
Heart Rate Source: ECG
The Panorama Central Station displays ECG waveforms when using 3-wire, 5-wire, and 12 
lead cable. The ECG leads available at the Panorama Central Station are dependent on the 
lead cable connected to the patient. 
Heart Rate Source: IBP 
12.1.2
Respiration: Resp
The Respiration parameter obtains data from ECG, CO
or Gas Module sources.
 
Unit of Measure:
Beats Per Minute (bpm)
Resolution:
1 bpm 
PATIENT SIZE
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
LEAD
WAVEFORM LEADS AVAILABLE
3-wire lead set
I, II, III
5-wire lead set
I, II, III, aVR, aVL, aVF, V
View 12
 card
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Unit of Measure:
mV
PATIENT SIZE
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
Unit of Measure:
Respirations Per Minute (rpm)
Resolution:
1 rpm