Mindray DS USA Inc. WT0707-00000 Manual De Usuario
Parameter Specifications
Parameter Specifications
12 - 2
0070-10-0690-01
Panorama™ Operating Instructions
12.1
Parameter Specifications
The parameters displayed on the Panorama Central Station are dependent on the type of
equipment connected to the Panorama system.
equipment connected to the Panorama system.
12.1.1
Heart Rate: HR
The Heart Rate parameter obtains data from ECG and IBP sources.
Heart Rate Source: ECG
The Panorama Central Station displays ECG waveforms when using 3-wire, 5-wire, and 12
lead cable. The ECG leads available at the Panorama Central Station are dependent on the
lead cable connected to the patient.
lead cable. The ECG leads available at the Panorama Central Station are dependent on the
lead cable connected to the patient.
Heart Rate Source: IBP
12.1.2
Respiration: Resp
The Respiration parameter obtains data from ECG, CO
2
or Gas Module sources.
Unit of Measure:
Beats Per Minute (bpm)
Resolution:
1 bpm
PATIENT SIZE
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
LEAD
WAVEFORM LEADS AVAILABLE
3-wire lead set
I, II, III
5-wire lead set
I, II, III, aVR, aVL, aVF, V
View 12
™
card
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Unit of Measure:
mV
PATIENT SIZE
MINIMUM VALUE (bpm)
MAXIMUM VALUE (bpm)
Neonate
30
350
Pediatric
30
300
Adult
30
300
Unit of Measure:
Respirations Per Minute (rpm)
Resolution:
1 rpm