Panorama™ Operating Instructions
0070-10-0690-01
12 - 5
Parameter Specifications
Parameter Specifications
12.1.5
Nonin SpO
2
The Nonin SpO
2
 parameter includes measurements from Nonin sensors.
The Panorama Central Station supports SpO
2
 waveforms.
12.1.6
IBP: mmHg
The Panorama Central Station supports up to four invasive blood pressure parameters. The 
measurements shown for each pressure include Systolic (Sys), Diastolic (Dia) and Mean.
NOTE:
IBP parameter labels include IBP, Art, UA, LV, PA, 
CVP, ICP, LA, and RA. The label used is based on the 
label selected at the bedside monitor.
IBP (Sys)
IBP (Dia) 
IBP Mean 
The Panorama Central Station supports IBP waveforms.
 
The Panorama Central Station will 
not display an IBP waveform if the IBP cable is not connected to the system. 
Unit of Measure:
%
Resolution:
1% 
PATIENT SIZE
MINIMUM VALUE (%)
MAXIMUM VALUE (%)
Neonate
N/A
N/A
Pediatric
N/A
N/A
Adult
---
100
Resolution:
1 mmHg
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
-30 300
Pediatric
-30 300
Adult
-30 300
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
-30 300
Pediatric
-30 300
Adult
-30 300
PATIENT SIZE
MINIMUM VALUE (mmHg)
MAXIMUM VALUE (mmHg)
Neonate
-30 300
Pediatric
-30 300
Adult
-30 300