Mindray DS USA Inc. WT0707-00000 Manual De Usuario
Panorama™ Operati
ng Ins
tructions
0070-10-0690-01
13 - 3
Alar
ms and Ev
en
ts
Physiological Alar
ms and Ev
en
ts
ALARM
ALARM
PRIORITY
ALARM LIMIT
NEONATE
PEDIATRIC
ADULT
DEVICES
Heart Rate Units: bpm
1 or 2
Factory Default: 2
Factory Default: 2
High Limit
Off, 100-350
Factory Default: 200
Factory Default: 200
Off, 100-300
Factory Default:180
Factory Default:180
Off, 60-250
Factory Default:150
Factory Default:150
Passport 2
®
Spectrum
™
/Spectrum OR
™
Telepack
Low Limit
Off, 30-200
Factory Default: 100
Factory Default: 100
Off, 30-150
Factory Default: 80
Factory Default: 80
Off, 30-120
Factory Default: 50
Factory Default: 50
Heart Rate Units: bpm
1 or 2
Factory Default: 2
Factory Default: 2
High Limit
N/A
N/A
Off, 60-250
Factory Default:150
Factory Default:150
Telepack with Nonin SpO
2
module
Low Limit
N/A
N/A
Off, 30-120
Factory Default: 50
Factory Default: 50
NOTE:
When the central station language is configured for French, the
Heart Rate (HR) high/low alarm limit cannot be turned Off.
Heart Rate (HR) high/low alarm limit cannot be turned Off.
NOTE:
When the selected device is a Telepack, the Low Heart Rate alarm
will be called when the heart rate becomes invalid because the rate
is underrange.
will be called when the heart rate becomes invalid because the rate
is underrange.
Respiration
Units: rpm
Units: rpm
1, 2, and 3
Factory Default: 3
Factory Default: 3
High Limit
Off, 30-200
Factory Default: Off
Factory Default: Off
Off, 15-150
Factory Default: Off
Factory Default: Off
Off, 10-100
Factory Default: Off
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Spectrum/Spectrum OR
Low Limit
Off, 5-50
Factory Default: 5
Factory Default: 5
Off, 5-40
Factory Default: Off
Factory Default: Off
Off, 5-30
Factory Default: Off
Factory Default: Off
SpO
2
Units: %
1 and 2
Factory Default: 2
Factory Default: 2
High Limit
Off, 80-100%
Factory Default: Off
Factory Default: Off
Off, 80-100%
Factory Default: Off
Factory Default: Off
Off, 80-100%
Factory Default: Off
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Spectrum/Spectrum OR
Low Limit
50-99%
Factory Default: 92%
Factory Default: 92%
50-99%
Factory Default: 90%
Factory Default: 90%
50-99%
Factory Default: 85%
Factory Default: 85%
SpO
2
(Nonin)
Units: %
1 and 2
Factory Default: 2
Factory Default: 2
High Limit
N/A
N/A
Off, 80-100%
Factory Default: Off
Factory Default: Off
Telepack with Nonin SpO
2
module
Low Limit
N/A
N/A
50-99%
Factory Default: 85%
Factory Default: 85%
NOTE:
The SpO
2
high/low alarm limit will always be displayed in the
digital tile portion of the Bedside tab All Waveforms view.
IBPx Sys.
Units: mmHg*
Units: mmHg*
2 or 3
Factory Default: 3
Factory Default: 3
High Limit
Off, 5-180
Factory Default: Off
Factory Default: Off
Off, 5-240
Factory Default: Off
Factory Default: Off
Off, 5-300
Factory Default: Off
Factory Default: Off
Passport 2
Spectrum/Spectrum OR
Spectrum/Spectrum OR
Low Limit
Off, 0-130
Factory Default: Off
Factory Default: Off
Off, 0-130
Factory Default: Off
Factory Default: Off
Off, 0-150
Factory Default: Off
Factory Default: Off
N/A - Not Applicable