Level
Immunity Test
Conducted RF
EN/IEC
61000-4-6
Radiated RF
EN/IEC
61000-4-3
EN/IEC 60601
Test Level
3 Vrms
150 kHz
to 80 MHz
10 V/m
80 MHz
to 2.5 GHz
10V/m
Compliance Electromagnetic
3 Vrms
*
Table 4 - Guidance and Manufacturer’s declaration – Immunity
The GoSafe Lite 7150MHB is intended for use in the electromagnetic
environment specified below. The customer or user of the GoSafe Lite
should ensure that it is used in such an environment.
The GoSafe Lite 7150MHB is intended for use in the electromagnetic
environment specified below. The customer or user of the GoSafe Lite
should ensure that it is used in such an environment.
Environment – Guidance
D=1.17√P
D=0.35√P
80 to 800 MHz
D=0.35√P
80 to 800 MHz
D=0.70√P
800 MHz to 2.5 GHz
Portable and mobile communications
Portable and mobile communications
equipment should be separated from
the GoSafe Lite 7150MHB by no less
than the distances calculated/ listed
below
*
:
D=(3.5/V1)(√P)
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
D=(3.5/E1)(√P)
80 to 800 MHz
D=(7/Eq)(√P)
800 MHz to 2.5 GHz
where P is the maximum power in
watts (W) and D is the recommended
separation distance in meters (m).
Field strengths from fixed RF
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,(a) should
be less than the compliance level in
each frequency range.(b)
Interference may occur in the vicinity
Interference may occur in the vicinity
of equipment marked with the
following symbol:
* NOTE:
Conducted RF Immunity does not apply to an internal battery operated
devices. Limit above is set for battery charger