Biolog-id SSTABIOLOGID Manual De Usuario

Descargar
Página de 27
 
Smart Storage Agitator User Manual
 
Version: 1.0 
 
Validation date of the manual: 18/09/2018                                                                                                    
 
   20 of 27
 
Guidance and manufacturer's declaration - electromagnetic immunity 
The PRD-7150300A is intended for use in the electromagnetic environment specified below. The customer or user of the 
PRD-7150300A should ensure that it is used in such an environment. 
Immunity test 
Test level 
According to IEC 
60601 
Level of  
conformity 
electromagnetic environment - guidelines 
 
 
 
 
 
 
 
 
 
Conducted RF 
disturbances 
IEC 61000-4-6 
 
 
 
Radiated RF 
disturbances 
IEC 61000-4-3 
 
 
 
 
 
 
 
 
3 Vrms 
150 kHz to 80 MHz 
 
 
 
 
3 V / m 
80 MHz to 2.5 GHz 
 
 
 
 
 
 
 
 
3 V 
 
 
 
 
 
 
3 V / m 
Portable and mobile RF communications devices should 
not be used closer to any part of the PRD-7150300A, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
 
Recommended separation distance 
d = 1.17 √
 
P 
 
d = √ P 1.17     80 MHz to 800 MHz 
 
d = √ P 2.34
   800 MHz to 2.5 GHz 
 
where P is the maximum output power characteristic of 
the transmitter in watts (W), according to the 
manufacturer of the transmitter and d is the 
recommended separation distance in meters (m). 
The field strengths of fixed RF transmitters determined by 
electromagnetic field investigation ᵃ should be less than 
the compliance level in each frequency range. ᵇ 
Interference may occur near the device marked with the  
 following symbol: 
 
NOTE 1 At 80 MHz and 800 MHz, the highest frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by the absorption and 
reflection of structures, objects and people. 
ᵃ The field strengths of fixed transmitters, such as base stations for radiotelephones (cellular / wireless) and land mobile 
radios, amateur radio, AM and FM broadcasting, and TV broadcasting, cannot be theoretically predicted accurately. To 
evaluate the electromagnetic environment due to fixed RF transmitters, an on-site electromagnetic investigation should be 
considered. If the field strength, measured at the location where the PRD-7150300A is used, exceeds the applicable RF 
compliance level above, the PRD-7150300A should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be required, such as reorienting or repositioning the PRD-7150300A. 
ᵇ Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V / m.