17 
 
Figure 5. OAS Pump control panel 
______________________________________________________________________ 
 
 
 
A.  Low saline red LED indicator 
B.  Prime button 
C.  Start button and green LED indicator 
D.  Status yellow LED indicator 
______________________________________________________________________ 
 
2.  Prepare a full 1000 mL bag of normal saline solution with lubricant.  Refer to the 
VIPERSLIDE Lubricant Instructions for Use for lubricant preparation instructions. 
3.  Hang the prepared saline bag with lubricant from the low saline level sensor on the 
standard IV pole. 
 
Caution: Do not use glass bottles for the saline solution with lubricant or hang 
multiple saline bags from the low saline level sensor as this will disable the Low Saline 
Information signal. 
 
11.4. OAD Set Up 
Carefully remove the OAD (or individual OAD handle and cartridge) from the tray(s) and 
set onto a stable surface.   
Note:  Removal of the OAD from the packaging by tipping the tray and allowing the OAD 
to fall out can result in damage to the OAD or ancillary devices underneath the OAD. 
11.4.1.  Connect the Cartridge to the Handle