Guangzhou Longest Science & Technology Co. Ltd. LGT-232 Manual De Usuario

Descargar
Página de 58
Troubles 
Possible causes 
Solutions 
3.  The status indicator 
on the unit flashed 
Yellow  and the unit 
switched off. 
The battery of the unit is empty. 
Charge the unit (see chapter 
Preparing for use’). 
4. There is not output. 
1. Did not press the start button; 
2.  Did not adjust the output 
intensity. 
1. Press the start button; 
2. Adjust the output intensity. 
5. Stimulation is 
uncomfortable. 
1.  The output intensity is too 
high; 
2.  Electrodes are too close 
together; 
3. Damaged or worn electrodes; 
4.  Electrode active area  size is 
too small. 
1. Decrease the output intensity; 
2. Reposition the electrodes; 
3. Replace; 
4. Replace electrodes with ones 
that have an active area no less 
than 25.0cm
2
6. The MStim Sport APP 
on the mobile phone pop 
up  tips  “electrodes fall 
off”, the indictor of the 
unit flashed Yellow. 
1.  The  electrodes are poor 
contact with the skin. 
2. Electrodes do not stick well; 
1. Re-connect the electrodes; 
2.  Replace electrodes. Apply 
electrodes to a clean,  dry 
surface.
 
7.  The unit  has output, 
but  treatment without 
sensation. 
The output intensity is too low. 
Increase the output intensity. 
FOR-MRRD0001-EN232 UM0.1