Guidance and manufacture’s declaration – electromagnetic immunity 
The LGT-235 is intended for use in the electromagnetic environment specified below. The 
customer or the user of LGT-235 should assure that it is used in such an environment. 
Immunity test 
IEC 60601 test 
level 
Compliance level 
Electromagnetic environment 
- guidance 
Electrostatic 
discharge 
(ESD) 
IEC 61000-4-2 
±6 kV contact 
±8 kV air 
±6 kV contact 
±8 kV air 
Floors should be wood, concrete 
or  ceramic  tile.  If  floor  are 
covered  with  synthetic  material, 
the relative humidity should be at 
least 30%. 
Electrical fast 
transient/ burst 
IEC 61000-4-4 
±2 kV for power 
supply lines 
±1 kV for 
input/output    lines 
±2kV for power 
supply lines 
Mains  power  quality  should  be 
that  of  a  typical  commercial  or 
hospital environment. 
Surge 
IEC 61000-4-5 
± 1 kV line(s) to 
line(s) 
± 2 kV line(s) to 
earth 
±1 kV differential 
mode 
Mains  power  quality  should  be 
that  of  a  typical  commercial  or 
hospital environment. 
Voltage dips, 
short 
interruptions 
and voltage 
<5% U
T
 (>95% dip 
in U
T
) for 0.5 cycle 
40% U
T
 (60% dip in 
<5% U
T
 (>95% dip 
in U
T
) for 0.5 cycle 
40% U
T
 (60% dip in 
Mains  power  quality  should  be 
that  of  a  typical  commercial  or 
hospital environment. If the user 
of 
the 
LGT-235 
requires