FOR-MRRD0001-EN235 UM0.1 
Troubles 
Possible causes 
Solutions 
3.  The  status  indicator 
on 
the 
unit 
flashed 
Yellow  and  the  unit 
switched off. 
The battery of the unit is empty. 
Charge  the  unit  (see  chapter 
Preparing 
for use’). 
4. There is not output. 
1. Did not press the start button; 
2.  Did  not  adjust  the  output 
intensity; 
3.  The  electrode  lead  hose 
connected badly; 
4.  The  electrode  lead  hose  has 
been damaged. 
1. Press the start button; 
2. Adjust the output intensity; 
3. Re-connect the electrode lead 
hose; 
4. Replace a new electrode lead 
hose. 
5. 
Stimulation 
is 
uncomfortable. 
1.  The  output  intensity  is  too 
high; 
2.  Damaged  or  worn  knee  pad 
for  electrotherapy  or  electrode 
lead hoses; 
3.  Electrode  active  area  size  is 
too small. 
1. Decrease the output intensity; 
2. Replace; 
3. Replace electrodes  with ones 
that have an active area no less 
than 25.0cm
2