User Manual for All-in-One Health Monitor 
 
 
Page XXVIII/69 
Between the left edge of the breast bone and Rib 5 
L(Yellow)/ 
LA(Black) 
F (Green)/ 
LL(Red) 
F (Green)/ 
LL(Red) 
2.4 Pressure Accuracy Verification 
Operation procedure: 
1. Unscrew the M3x6 screw from the battery compartment on the back of the All-in-One Health Monitor, as shown in 
figure 2.9. 
2. Take a Cuff connector plug from the battery cover, as shown in figure 2.10. (Note: there are two plugs but you will 
only need one.)   
3. Air Path Connection: Take a piece of air tube (0.5~1m long, Φ8.0mm/Φ4.0mm diameter). Attach the Cuff connector 
with a connector plug on to one end of the air tube. Connect the other end to the 3-way connector. Connect the 
other 2 ends of the 3-way connector to a pressure meter such as Mercury Sphygmomanometer as shown in Figure 
2.12. 
4. Connect the Cuff connector end to the NIBP port on the All-in-One Health Monitor.