7.1.2 Routine Maintenance
The All-in-One Health Monitor is designed to have a life expectancy of at least 5 years. In order to ensure precise
readings, it is recommended to test the pressure accuracy every year. Please see section 2.4 to perform the Blood
Pressure Accuracy Check Method or contact your supplier for more information. After each maintenance or yearly
maintenance, the monitor should be thoroughly inspected by a qualified personnel, including an inventory of all
functions and safety examinations.
readings, it is recommended to test the pressure accuracy every year. Please see section 2.4 to perform the Blood
Pressure Accuracy Check Method or contact your supplier for more information. After each maintenance or yearly
maintenance, the monitor should be thoroughly inspected by a qualified personnel, including an inventory of all
functions and safety examinations.
If the hospital fails to carry out a satisfactory maintenance program on the monitor, it may cause harm to the patient.
If there is any indication of cable and transducer deterioration or damage, please do not use.
The SpO
2
function has been calibrated before vending. If the user needs to verify the precision of SpO
2
function, the
simulator with model FLUKE INDEX2 can be used.
Any adjustable unit or component inside the monitor cannot be adjusted without permission so as to avoid
unnecessary failures that may affect the normal application.
unnecessary failures that may affect the normal application.
It is recommended to use the battery once a month to ensure the battery's normal capacity and its long service
life, and recharge once the battery has completely run out.
life, and recharge once the battery has completely run out.