User Manual for All-in-One Health Monitor 
 
 
Page L/69 
7.1.2 Routine Maintenance 
The All-in-One Health Monitor is designed to have a life expectancy of at least 5 years. In order to ensure precise 
readings, it is recommended to test the pressure accuracy every year. Please see section 2.4 to perform the Blood 
Pressure Accuracy Check Method or contact your supplier for more information. After each maintenance or yearly 
maintenance, the monitor should be thoroughly inspected by a qualified personnel, including an inventory of all 
functions and safety examinations.   
 
If the hospital fails to carry out a satisfactory maintenance program on the monitor, it may cause harm to the patient. 
 
If there is any indication of cable and transducer deterioration or damage, please do not use. 
 
The SpO
function has been calibrated before vending. If the user needs to verify the precision of SpO
function, the 
simulator with model FLUKE INDEX2 can be used. 
 
Any  adjustable unit or component inside the monitor cannot be adjusted without permission so as to avoid 
unnecessary failures that may affect the normal application. 
 
It is recommended to use the battery once a month to ensure the battery's normal capacity and its long service 
life, and recharge once the battery has completely run out.