Activating Your Device 
 
The ActivePERS devices will arrive in a mode designed to save power while sitting on a 
shelf.  In this mode, the sensors will not detect falls or send messages of any kind.   
 
To exit this mode and enter active mode, press and hold the button until the LED turns 
on, and then release the button.  This will take three seconds.  Once this occurs, the 
device will operate in its normal mode.  
 
NOTE: This is a one-time process and cannot be reversed. (i.e. the device cannot go 
back to shelf-mode after activation.) 
 
Detecting a Fall 
 
In the case of a fall detected automatically, the LED will turn on, and a fall alert 
message will be transmitted. 
 
Measuring Heartbeats 
 
The devices send heartbeat messages once per hour.  These messages contain the 
accumulated activity information for that time period, with values determined by the 
number of steps taken divided by 10.  For example, if the user took 530 steps in the 
time period, the activity level will be reported ad 53.   
The heartbeat message can be identified by the only signal where both the fall and 
panic bits are set to zero with the activity level below 255.  The activity level is reset 
after every heartbeat message. 
Sending Panic Signal 
 
To send a panic signal, press and release the button.  The LED should turn on upon 
release, and a message will be transmitted to indicate the panic.  
 
Clearing Alarm 
 
To clear an alarm and turn off the LED, press and hold the button until the LED 
turns off, which will take approximately 10 seconds.  This will transmit a clear 
alarm message, indicated by the activity level being set to 255. 
 
The LED will turn on and remain on after sending a fall or a panic signal.  
Sending a clear alarm signal will turn the LED off.