14.2 Environmental Specifications
Operating Conditions
LuViva
CG
Temperature
10
C - 40
C
(50
F – 104
F)
10
C - 40
C
(50
F – 104
F)
Humidity
30% to 75%
non-condensing
30% to 75%
non-condensing
Altitude
700 hPa to 1060 hPa
700 hPa to 1060 hPa
Transport Conditions – Within
Same Facility
LuViva
CG
Temperature
10
C - 40
C
(50
F – 104
F)
10
C - 40
C
(50
F – 104
F)
Humidity
10% to 95%
non-condensing
10% to 95%
non-condensing
Altitude
700 hPa to 1060 hPa
700 hPa to 1060 hPa
Transport Conditions – Between
Different Facilities
LuViva
CG
Temperature
-20
C to 50
C
(-4
F – 140
F)
-20
C to 50
C
(-4
F – 140
F)
Do not expose cervical
guides to temperatures
guides to temperatures
above 50C (122F). Guided
Therapeutics, Inc. cannot
guarantee the performance
guarantee the performance
of the cervical guide when
stored at temperatures
outside of the recommended
outside of the recommended
range.
Humidity
10% to 95%
non-condensing
10% to 95%
non-condensing
Special Instruction
Transport in packaged
configuration
CG-20 packaged
configuration
Storage Conditions
LuViva
CG
Temperature
-20
C to 50
C
(-4
F to 140
F)
The cervical guides are best
when stored at
-20
C to 30
C
(-4
F to 86
F)
Humidity
10% to 95%
non-condensing
10% to 95%
non-condensing