Centauri Medical, Inc.  
 
  Document No. 1312AJ 
 
DynaSense Instructions For Use
 
 
Page 13 of 54
 
The Patient Sensor should be removed if current is going to be passed 
through the patient’s skin or body, such as during external cardiac 
defibrillation or electrocautery. In the event that the sensor is not removed 
and the subject has a medical emergency that requires external cardiac 
defibrillation or exposure to high frequency surgical equipment, the subject 
may experience a skin burn where the sensor has been applied. 
 
If a patient receives an MRI study, the Patient Sensor should be removed. 
If the sensor is not removed, the patient may experience tugging from the 
MRI magnet or skin heating where the sensor has been applied. 
 
DynaSense has not been tested for compatibility on patients with 
pacemakers or internal cardiac defibrillators. As such, patients with 
pacemakers or internal cardiac defibrillators should not use the Patient 
Sensor. Electrical interference to a pacemaker or internal cardiac 
defibrillator from the Patient Sensor is unlikely, but has not been tested. 
 
 
Other wireless communications may interfere with the proper functioning of 
DynaSense, or conversely DynaSense may interfere with the proper 
functioning of other wireless communication systems.  However, the risk of 
interference has been mitigated. DynaSense is designed so that incidental 
interference will not put the patient at significant risk.  Installation is 
required to be completed by qualified personnel, who can be contacted via 
Customer Service, to further minimize interference with 802.11 (or WiFi) 
networks.  If the 802.11 networks undergo configuration changes or if new 
networks are being configured in the 2.4GHz spectrum, please contact 
Customer Service. 
 
 
BENEFITS 
 
The Patient Sensor monitors a patient’s orientation and movements and 
communicates this data wirelessly to the User Interface.  From the User