Warning: Avoid using insulin from more than one vial, which may introduce air into the syringe. 
8.  With the vial down, hold the syringe. Pull straight up to remove the syringe needle from the vial, and then insert it 
straight down into the insulin fill port on the underside of the Reservoir Patch. 
 
Warning: To ensure proper fill, do NOT insert fill syringe at an angle into the fill port. 
Warning: Only use specified needle and syringe. 
9.  Depress the syringe plunger to completely empty the insulin into the Reservoir Patch. The PDM will beep and/or 
vibrate, indicating that the System is ready to proceed to the next step. 
Warning:  NEVER use a Reservoir Patch if you hear a crackling noise or feel resistance when you depress the 
plunger. These conditions can result in under delivery of insulin. 
10.  Remove the needle from the insulin fill port. The port is self-sealing; insulin will not leak after the needle is 
removed. 
Warning: Do NOT insert the fill syringe into the fill port more than once. 
Warning: NEVER inject air into the fill port. Doing so may result in unintended or interrupted insulin delivery. 
11.  Use your reservoir right after you fill it. Do NOT store it filled. 
12.  Place the protective cap back on the needle and remove the needle from the syringe. 
13.  Safely dispose of the capped needle in a sharps container according to local waste disposal regulations. 
Warning: The fill syringe is intended for single use only, and only specified fill syringe can be used.