months when it is no regular used. It can extend the battery life following this guidance. 
D.Users are advised to calibrate the device termly (or according to the calibrating program of 
hospital). It also can be performed at the state-appointed agent or just contact us for calibration. 
7.3.  Transportation and storage 
A. The packed device can be transported by ordinary conveyance or according to transport contract.     
The device can not be transported mixed with toxic, harmful, corrosive material. 
B. The packed device should be stored in room with no corrosive gases and good ventilation. 
Temperature: -40°C~60°C; Relative Humidity: ≤95% 
8.  Troubleshooting
 
Trouble 
Possible Reason 
Solution 
The SpO
and Pulse 
Rate can not be 
displayed normally 
1. The finger is not properly positioned. 
2. The patient’s SpO
is too low to be 
detected.  
1. Lay the finger properly and try 
again.  
2. Try again; Go to a hospital for 
a diagnosis if you are sure the 
device works all right. 
The SpO
and Pulse 
Rate are not 
displayed stably 
1. The finger is not laid inside deep 
enough. 
2. The finger is shaking or the patient is 
moving. 
1. Lay the finger properly and try 
again.  
2. Let the patient keep calm   
The device can not 
be turned on 
1. The batteries are drained or almost 
drained. 
2. The device’s malfunction 
1.Please recharge the battery 
2. Please contact the local 
service center. 
The display is off 
suddenly 
1. This device is set to be automatically 
power off within 5 seconds when it cannot 
detect any signal (without external probe) 
2. The battery is drained away or almost 
drained away . 
1.Normal 
2.Please recharge the battery 
The battery can not 
be full charged even 
after 10 hours 
charging time. 
  The battery is broken 
Please contact the local service 
center. 