RF emission 
CISPR 11 
Class B 
The 
CMS50EW 
is suitable for use in all establishments, 
including domestic and those directly connected to a low 
voltage power supply network which supplies buildings 
used for domestic purposes. 
 
 
Guidance and manufacturer’s declaration – electromagnetic immunity – 
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING 
Guidance and manufacturer’s declaration – electromagnetic immunity 
The  CMS50EW  is intended for use in the electromagnetic environment specified below. The customer or the user of CMS50EW 
should assure that it is used in such an environment. 
Immunity test 
IEC 60601 test level 
Compliance level 
Electromagnetic environment - 
guidance 
Electrostatic 
discharge (ESD) 
IEC 61000-4-2 
±6 kV contact 
±8 kV air 
±6 kV contact 
±8 kV air 
Floors should be wood, concrete or 
ceramic tile. If floor are covered 
with synthetic material, the relative 
humidity should be at least 30%. 
Power frequency 
(50/60Hz) magnetic 
field 
IEC 61000-4-8 
3A/m 
3A/m 
Mains power quality should be that 
of a typical commercial or hospital 
environment. 
NOTE    
Guidance and manufacturer’s declaration – electromagnetic immunity
 
The  CMS50EW  is intended for use in the electromagnetic environment specified below. The customer or the user of CMS50EW 
should assure that it is used in such an environment.
 
Immunity test 
IEC 60601 test level 
Compliance 
level 
Electromagnetic environment - guidance 
 
 
 
 
 
 
 
Conducted RF 
IEC 61000-4-6 
 
 
Radiated RF 
IEC 61000-4-3 
 
 
 
 
 
 
 
 
 
3 V
rms
 
150 kHz to 80 MHz 
 
 
3 V/m 
80 MHz to 2.5 GHz 
 
 
 
 
 
 
 
 
3 V
rms
 
 
 
 
3 V/m 
Portable and mobile RF communications equipment should 
be used no closer to any part of the CMS50EW, including 
cables, than the recommended separation distance 
calculated from the equation applicable to the frequency of 
the transmitter. 
Recommended separation distance 
P
V
d
=
1
5
.
3
 
 
P
E
d
=
1
5
.
3
 
80 MHz to 800 MHz 
P
E
d
=
1
7
 
800 MHz to 2.5 GHz