Guidance and manufacturer’s declaration – electromagnetic emissions
 
The “CMS50K” is intended for use in the electromagnetic environment specified below. The customer or the 
user of the “CMS50K” should ensure that it is used in such an environment. 
Emissions test 
Compliance 
Electromagnetic environment – guidance 
RF emissions 
CISPR 11
 
Group 1
 
The “CMS50K” uses RF energy only for its internal 
function. Therefore, its RF emissions are very low and are 
not likely to cause any interference in nearby electronic 
equipment. 
RF emissions 
CISPR 11
 
Class B
 
The “CMS50K” is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply network 
that supplies buildings used for domestic purposes. 
Harmonic emissions 
IEC 61000-3-2
 
Class A
 
 
Voltage fluctuations/ 
flicker emissions 
IEC 61000-3-3
 
Complies 
 
 
 
Recommended separation distances between 
portable and mobile RF communications equipment and the Medical CMS50K
 
The “CMS50K” is intended for use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the Medical CMS50K can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) 
and the “CMS50K” as recommended below, according to the maximum output power of the communications 
equipment. 
Rated maximum output 
power of transmitter 
W 
Separation distance according to frequency of transmitter 
m 
 
150 kHz to 80 MHz 
 
 
80 MHz to 800 MHz 
 
 
800 MHz to 2,5 
GHz 
 
 
 
0,01 
0.12 
0.12 0.23 
0,1 
0.38 
0.38 0.73 
1.2 
1.2 2.3 
10 
3.8 
3.8 7.3 
100
 
12 
12 23 
For transmitters rated at a maximum output power not listed above, the recommended separation distance in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where is the 
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. 
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people.