Contec Medical Systems Co. Ltd. PM10 Manual De Usuario

Descargar
Página de 12
 
 
Chapter 3 Primary Technical Orders 
3.1 Normal work environment 
1)  Operation environment 
  Temperature: +5 ~+40
 
  Relative humidity: 25%~80% 
  Atmospheric pressure:70kPa~106kPa 
  Power supply: built-in rechargeable lithium battery, voltage: 3.7V 
2)  Transportation and storage environment 
  Temperature: -40℃~+55℃ 
  Relative humidity: ≤95% 
  Atmospheric pressure:50kPa~106kPa 
3.2 Basic parameters 
1)  Calibration voltage: 1mV±5% 
2)  Standard sensitivity: 10mm/mV±5% 
3)  Amplitude frequency characteristic: standard: 10Hz; 1Hz~20Hz; (+0.4dB,-3dB) 
4)  Noise level: ≤30μV 
5)  CMRR: ≥60dB 
6)  Scanning speed: 25mm/s±5% 
7)  Sampling rate: 250 dots/s 
8)  HR measurement range: 30bpm~300bpm, error:  ±1bpm or 1% 
9)  Battery Voltage: DC 3.7V 530mAh 
10)  Type of protection against electric shock: Internal power device 
11)  Degree of protection against electric shock: Type BF applied part 
12)  Degree of waterproof: IP22