Clean the UltraMIST System prior to and after each treatment session with the Power Switch in the “off” position.
CLEANING THE ULTRAMIST SYSTEM
Do not sterilize the treatment wand including tip with steam,
ETO, radiation, gas/plasma, or cold sterilants. After completion
of the UltraMIST treatment, repeat the cleaning process.
PREPARING THE PATIENT FOR TREATMENT WITH THE ULTRAMIST SYSTEM
Collect all required materials (generator, treatment wand,
applicator, IV saline, germicidial wipes, absorbent pads
and gloves).
Use facility approved Infection Control Procedures
throughout the treatment session.
Treatment with the UltraMIST System will require a clean
environment.
Depending on the wound location, position the patient in
a comfortable position for treatment.
WARNING: The UltraMIST applicator is designed as a single patient-use disposable unit to avoid contamination. Do not
re-sterilize or reuse applicators. Re-using the applicator and/or saline may result in infection and degraded performance.
DESCRIPTION & INDICATIONS FOR USE
Description: The UltraMIST
TM
System delivers low frequency ultrasound to the treatment site using a noncontact saline mist.
Indications for Use: The MIST Therapy System produces a low energy ultrasound-generated mist used to promote wound healing
through wound cleansing and maintenance debridement by the removal of fibrin, yellow slough, tissue exudates and bacteria.
Contraindications: Do not use near electronic implants/prosthesis (e.g. Near or over the heart or over the thoracic area if the patient
is using a cardiac pacemaker); on the lower back during pregnancy or over the pregnant uterus; over areas of malignancies.
Potential Complications: Tingling, redness.
Potential Complications: Tingling, redness.
CONTRAINDICATIONS & POTENTIAL COMPLICATIONS
Prior to treatment, clean the entire system using facility approved
infection control procedures.
PL-61044_B
2
1
2
3
4
Remove soiled dressings and discard all contaminated
materials (gloves, dressings, pads) prior to UltraMIST
treatment.
Place an absorbent pad under the portion of the patient
being treated to collect the saline runoff.
Measure the area of the wound to be treated.
5
6
7
1
2
ULTRAMIST™
SYSTEM
INSTRUCTIONS FOR USE