Bruker Optik GmbH BRAVO-01 Manual De Usuario
31
BRAVO Analyzer - User Manual
Operation 4
4.5
Verifying a sample
1. Putting the sample in the measurement
position
In case you verify samples which are filled in
vials, place the vial in the holder of the mea-
suring tip.
vials, place the vial in the holder of the mea-
suring tip.
In case you verify samples which are filled in
bags, press the bag against the laser exit
opening of the measuring tip.
bags, press the bag against the laser exit
opening of the measuring tip.
☞ In case the measuring tip needs to be
exchanged, see section 4.4.
2. Enabling the laser (only if not yet done)
Press the button
at the BRAVO button
panel. If the laser is already enabled, skip
this step.
this step.
i
An enabled laser is indicated by a blue
(solid) laser LED.
Note: When you start a measurement with a
disabled laser, BRAVO aborts immediately
the measurement and delivers a result of the
category Aborted. The reason stated for the
abortion is Laser not enabled.
disabled laser, BRAVO aborts immediately
the measurement and delivers a result of the
category Aborted. The reason stated for the
abortion is Laser not enabled.
3. Starting the sample measurement
Either press the button
at the BRAVO
button panel
...
... or touch the icon
in the navigation bar
of the Collect Data screen.
☞ For more details, refer to the BRAVO
Software Manual. This manual is deliv-
ered as pdf file on an USB flash drive.
ered as pdf file on an USB flash drive.
Draft version